Bad filling ionic liquid sample in split tube furnace
الناشر \ المحرر \ الكاتب (Author / Editor / Publisher)
علي جاسم حسين الزهيري
Citation Information
علي,جاسم,حسين,الزهيري ,Bad filling ionic liquid sample in split tube furnace , Time 13/01/2018 20:39:09 : كلية الهندسة/المسيب
وصف الابستركت (Abstract)
Vapors of sample crystallization water confined inside the quartz tube and spilled out toward the furnace flanges due to high temperature
الوصف الكامل (Full Abstract)
Completely filled crucible with ionic liquid (10ml) treated with high temperature (1000oC) causes sample surface to solidify. Water of crystallization starts to boil inside the crucible leads to sudden pressure explosion for the sample. The opened side valves protected the quartz tube from breakage Figure 6. Assuming that furnace valves were closed in case of vacuum or under certain gases, serious hazard will occur due to glass breakage. Lesson Learned In case of such salts (ionic liquids) thermal treatment, sample preparation must be done in appropriate amount to ensure uniform heat distribution to all parts of the sample which prevents drying some parts without the other as stated in the case provided. A successful experiment was done keeping in mind the lessons derived from past experience